مشخصات پژوهش

صفحه نخست /Ultra-sensitive determination ...
عنوان Ultra-sensitive determination of insulin on pencil graphite electrode modified by cerium salen encapsulated zeolite (CS@Z-PGE)
نوع پژوهش مقاله چاپ‌شده در مجلات علمی
کلیدواژه‌ها Insulin, Cerium Salen, Encapsulated, Zeolite, Pencil graphite electrode
چکیده In this study, cerium salen encapsulated into dealuminated zeolite Y (CS@Z) was prepared through template synthesis method. This heterogeneous catalyst was characterized by XRD, FT-IR, SEM and EDS analysis. The amount of CS in the nanocage of zeolite was obtained 0.0846g/g of support. Pencil graphite electrode modified by CS@Z was applied for electrochemical ultra-sensitive determination of insulin. Parameters influencing the method were optimized. A linear range calibration curve from 0.05-50 nM of insulin with a detection limit of [12.8 (±0.6) pM] (based on S/N=3) and sensitivity of [44.3 (±0.8) μA μM-1] was observed at pH 7.4 and applied potential of 0.45 V by amperometry. The results demonstrated that the sensor can be used for the determination of insulin in the commercially available injections.
پژوهشگران سیما پور بیرام (نفر اول)، مریم موسوی فر (نفر دوم)، لاله اشتری (نفر سوم)